Steam sterilization processes affect the stability of clinical thermometers: Thermistor and prototypal FBG probe comparison

2020

Abstract Temperature is one of the most frequently measured physical quantities in clinical environment and a good biomarker of illness. The need for reusable probes, which have to be sterilized to prevent infections, requires the metrological qualification of thermometer probes in response to ageing effects induced by several sterilization processes. In this study, we investigated the effect of repeated sterilizations on both a commonly-used autoclavable thermistor probe and a prototypal Fiber Bragg Gratings (FBG) probe for temperature measurements. Findings highlighted a greater reliability of the proposed FBG probe than the commercial thermistor. Specifically, the FBG probe was able to w…

Accuracy and precisionMaterials science02 engineering and technology01 natural sciencesTemperature measurement010309 optics020210 optoelectronics & photonicsFiber Bragg gratingThermistor0103 physical sciences0202 electrical engineering electronic engineering information engineeringBiomedical sensorsElectrical and Electronic EngineeringInstrumentationbusiness.industryThermistorSterilizationAtomic and Molecular Physics and OpticsSteam sterilizationElectronic Optical and Magnetic MaterialsFiber Bragg Grating (FBG)Control and Systems EngineeringThermometerBiomedical sensorOptoelectronicsClinical thermometerbusinessOptical Fiber Technology